Remote Control Commands
R&S
®
SMU200A
490Operating Manual 1007.9845.32 ─ 15
*CLS
CLear Status
Sets the status byte (STB), the standard event register (ESR) and the EVENt part of the
QUEStionable and the OPERation registers to zero. The command does not alter the
mask and transition parts of the registers. It clears the output buffer.
Setting onlyUsage:
*ESE <Value>
Event Status Enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
0 to 255
<Value>
Range:
*ESR?
Event Status Read
Returns the contents of the event status register in decimal form and subsequently sets
the register to zero.
Return values:
0 to 255
<Contents>
Range:
Query onlyUsage:
*IDN?
IDeNtification: returns the instrument identification.
Return values:
"Rohde&Schwarz,<device type>,<serial number>,<firmware
version>"
<ID>
Rohde&Schwarz,ZVA8-4Port,12345,0.10.1.23
Example:
Query onlyUsage:
Common Commands