EasyManua.ls Logo

Rohde & Schwarz SMU200A - Bert:setup:data[:Polarity]; Bert:setup:denable; Bler:setup:data[:Polarity]; Bler:setup:denable

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
517Operating Manual 1007.9845.32 ─ 15
Parameters:
LSB|MSB
LSB
The checksum starts with the least significant byte.
MSB
The checksum starts with the most significant byte.
LSB
<Corder>
*RST:
BLER:SET:CORD MSB
the checksum starts with the most significant byte.
Example:
:BLER:SETup:DATA[:POLarity] <Polarity>
:BERT:SETup:DATA[:POLarity] <Polarity>
Sets the polarity of the feedback data bits.
Parameters:
NORMal|INVerted
NORMal
High level stands for a logic 1, low level for a logic 0.
INVerted
Low level stands for a logic 1, high level for a logic 0.
NORMal
<Polarity>
*RST:
BERT:SET:DATA INV
sets inversion of data signal.
Example:
:BLER:SETup:DENable <Denable>
:BERT:SETup:DENable <Denable>
Activates/deactivates the use of the Data Enable signal and the polarity of the signal if
it is used. The Data Enable signal marks the data that is actually to be evaluated for the
BER measurement. Any data in addition to the PRBS sequence is masked and thus not
evaluated (e.g. sync, preambles, other channels, etc that are present in the data bits
supplied by the DUT).
BERT and BLER Subsystems

Table of Contents

Other manuals for Rohde & Schwarz SMU200A

Related product manuals