Remote Control Commands
R&S
®
SMU200A
560Operating Manual 1007.9845.32 ─ 15
Parameters:
AVERage|PEAK
AVERage
<Priority>
*RST:
SENS1:DISP:PERM:STAT ON
the permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
the measured average power is indicated.
Example:
SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
The command queries the current filter length for auto filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE AUTO)
Return values:
float<Auto>
SENS1:FILT:TYPE AUTO
selects auto filter mode for the power sensor connected to the
SENSOR connector.
SENS1:FILT:LENG:AUTO?
queries the automatically set filter length.
Response: 1024
Example:
Query onlyUsage:
SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
The command selects the filter length for user filter mode
(SENSe:POWer:FILTer:TYPE USER). As the filter length works as a multiplier for the
time window, a constant filter length results in a constant measurement time. Values 1
and 2^n are setable.
The time window is fixed to 20 ms.
Parameters:
float
1 to 65536
1
<User>
Range:
*RST:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:LENG 16
sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
Example:
SENSe, READ and INITiate Subsystems