*IST?
Individual STatus query
Returns the contents of the IST flag in decimal form. The IST flag is the status bit which
is sent during a parallel poll.
Return values:
0 | 1<ISTflag>
Query onlyUsage:
*OPC
OPeration Complete
Sets bit 0 in the event status register when all preceding commands have been executed.
This bit can be used to initiate a service request. The query form writes a "1" into the
output buffer as soon as all preceding commands have been executed. This is used for
command synchronization.
*OPT?
OPTion identification query
Queries the options included in the instrument. For a list of all available options and their
description refer to the CD-ROM.
Return values:
The query returns a list of options. The options are returned at
fixed positions in a comma-separated string. A zero is returned for
options that are not installed.
<Options>
Query onlyUsage:
*PCB <Address>
Pass Control Back
Indicates the controller address to which remote control is returned after termination of
the triggered action.
Setting parameters:
0 to 30
<Address>
Range:
Setting onlyUsage: