Instrument Function
R&S
®
SMU200A
180Operating Manual 1007.9845.32 ─ 15
If the data is not cyclically continued, the measurement can be selectively interrupted and
restarted. Certain data sections and frames that have been marked as faulty can be
excluded from the measurement.
5.4.10.1 General Settings
State
Switches the bit error rate measurement on/off. Depending on the selection in the
"Mode" box, either a continuous measurement ("Auto") or a single measurement ("Sin-
gle") is carried out. A single measurement must be started by clicking the "Execute Sin-
gle" button.
SCPI command:
:BERT:STATe on page 521
Bit Error Rate
Displays the measured bit error rate.
The bit error rate is the ratio of occurred errors to transmitted data bits.
The display unit can be selected in the box to the right of the value.
"ENG"
The bit error rate is output in exponential notation with the exponent –3
or –6.
"SCI"
The bit error rate is output in scientific notation, i.e. standardized to one
place to the left of the decimal.
"%"
The bit error rate is output in percent.
"ppm"
The bit error rate is output in parts per million.
SCPI command:
:BLER:RESult on page 513
:BERT:UNIT on page 522
Bit and Block Error Rate Measurements - BERT Block