Remote Control Commands
R&S
®
SMU200A
622Operating Manual 1007.9845.32 ─ 15
Parameters:
9|11|15|16|20|21|23
9
<Length>
*RST:
BB:DM:SOUR PRBS
The internal pseudo-random generator is used as the data source.
BB:DM:PRBS 9
An internal pseudo-random sequence of 511 bits will be gener-
ated.
Example:
[:SOURce<hw>]:BB:DM:PRESet
The command calls the default settings for digital modulation.
BB:DM:PRES
calls the default settings for DM.
Example:
EventUsage:
[:SOURce<hw>]:BB:DM:SOURce <Source>
The command selects the data source.
Parameters:
ZERO|ONE|PRBS|PATTern|DLISt | SERial|PARallel
ZERO
An internally generated 0 data sequence is used.
ONE
An internally generated 1 data sequence is used.
PRBS
The pseudo-random sequence generator is used as the data
source.The length of the random sequence is defined with the aid
of command SOURce:BB:DM:PRBS.
PATTern
Internally generated data with a fixed pattern is used. The data
pattern is dafined using command SOURce:BB:DM:PATTern.
DLISt
Data from the selected data list is used.
SERial
External data from the serial interface is used.
In case of two-path instruments, this affects path A.
PARallel
External data from the serial interface is used.
In case of two-path instruments, this affects path A.
PRBS
<Source>
*RST:
SOURce Subsystem