Remote Control Commands
R&S
®
SMU200A
554Operating Manual 1007.9845.32 ─ 15
Parameters:
AUTO|FIXed | NORMal|HPOWer
AUTO
The electronically switching attenuator switches with a 5 dB step
width at fixed switching points.
With option High Power: The level settings are made in the area
of the electronically switching attenuator as well as the relay-
switched option. The entire level range is available.
FIXed
The attenuator is fixed at the current position. The uninterrupted
level settings are made if automatic level control is activated
(SOURce:POWer:ALC ON).
With option High Power: The level settings are made without
switching the high-power output option. When this operating mode
is switched on, the options are fixed in their current positions and
the resulting variation range is defined.
NORMal
(with Option High Power only)
The level settings are made only in the area of the electronically
switching attenuator. The high level ranges are not available.
HPOWer
(with Option High Power only)
The level settings are made only in the area of the option. Only the
high level range is available. The relays are not switched.
AUTO
<Amode>
*RST:
POW:ALC ON
activates automatic level control for RF output.
OUTP:AMOD FIX
sets the fixed mode with uninterrupted level for RF output.
Example:
:OUTPut<hw>:BLANk:LIST:STATe <State>
The command activates/deactivates RF output blanking.
Parameters:
0|1|OFF|ON
ON
<State>
*RST:
OUTP:BLAN:LIST:STAT ON
activates the RF Output blanking.
Example:
:OUTPut:BLANk:POLarity <Polarity>
The command sets the polarity of the No Signal (Blank) Marker.
OUTPut Subsystem