Remote Control Commands
R&S
®
SMU200A
519Operating Manual 1007.9845.32 ─ 15
:BERT:SETup:MCOunt <Mcount>
Enters the number of transmitted data bits/data blocks to be checked before the mea-
surement is terminated. With a BER measurement, data that was suppressed by
BERT:SETup:DENable is not counted. This termination criterion always terminates the
measurement after the specified number of data bits/data blocks. Starting from this point,
the fourth value is output with 1 (= terminate measurement) if the result is queried
with :BERT|BLER:RES. If the continuous measurement mode (BERT|BLER:SEQ
AUTO) has been selected, the measurement is restarted once the results have been
queried.
Parameters:
float
0.0 to 4294967295.0
BERT: 10000000 / BLER: 100000
<Mcount>
Range:
Default unit:
BERT:SET:MCO 1E6
sets a measurement over 1000000 data bits.
BLER:SET:MCO 1E3
sets a measurement over 1000 data blocks.
Example:
Example:
:BLER:SETup:MERRor <Merror>
:BERT:SETup:MERRor <Merror>
Enters the number of errors to occur before the measurement is terminated. This termi-
nation criterion always terminates the measurement after the specified number of errors.
Starting from this point, the fourth value is output with 1 (= terminate measurement) if the
measurement result is queried with :BERT|BLER:RES.
Parameters:
float
0.0 to 4294967295.0
100
<Merror>
Range:
*RST:
BERT:SET:MERR 1E4
sets a measurement over 1000 errors.
Example:
:BERT:SETup:RESTart:STATe <State>
Activates/deactivates an external restart of the BER measurement.
BERT and BLER Subsystems