Remote Control Commands
R&S
®
SMU200A
525Operating Manual 1007.9845.32 ─ 15
:CALibration<hw>:IQModulator:FULL?
The command starts adjustment of the I/Q modulator for the entire frequency range. The
I/Q modulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Return values:
0|1<Full>
CAL:IQM:FULL?
starts adjustments for the I/Q modulator across the entire fre-
quency range.
Response: "0"
adjustment has been performed successfully
Example:
Query onlyUsage:
:CALibration<hw>:IQModulator:LOCal?
The command starts adjustment of the I/Q modulator for the currently set frequency, I/Q
swap and baseband gain. The I/Q modulator is adjusted with respect to carrier leakage,
I/Q imbalance and quadrature.
Return values:
0|1<Local>
CAL:IQM:LOC?
starts adjustment for the I/Q modulator for the currently set fre-
quency.
Response: 0
adjustment has been performed successfully
Example:
Query onlyUsage:
:CALibration<hw>:LEVel[:MEASure]?
The command starts all adjustments which affect the level.
In case of two-path instruments, this affects the selected path.
Return values:
0|1<Measure>
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Example:
Query onlyUsage:
CALibration Subsystem