Instrument Function
R&S
®
SMU200A
230Operating Manual 1007.9845.32 ─ 15
Range
Sets the range for filling the table.
SCPI command:
n.a.
Select column to fill
Selects either the frequency or the level column to be filled with the value defined below.
SCPI command:
n.a.
Start value
Sets the start value for the frequency or the level entries.
SCPI command:
n.a.
End value
Displays the end value for the frequency or the level entries.
SCPI command:
n.a.
With increment
Sets the increment for the frequency or the level entries.
SCPI command:
n.a.
Fill
Fills the selected column in the set range with values, starting with the start value and
using the set increment.
SCPI command:
n.a.
5.5.9.3 Filling the Correction List with Power Sensor Measurement Data
The level correction values for the user correction list can be acquired by means of R&S
NRP power sensors. The R&S NRP sensors are connected to either the SENSOR con-
nector or to one of the USB interfaces. Configuration of the connection is performed in
the "Power Sensor" menu (see chapter 5.5.6, "NRP-Z Power Viewer", on page 210). The
filling of the ucor list with measurement data is performed in the ucor list editor (see "Edit
User Cor. Data - User Correction" on page 225).
In the editor, the frequencies for which the correction values are to be acquired are
entered in the frequency column (either manually or by means of the "Fill..." menu).
Do not save the list at this point, because the frequency entries are lost as long as there
are no entries for the level column also. In the following these entries are automatically
acquired by the connected power sensor.