Instrument Function
R&S
®
SMU200A
225Operating Manual 1007.9845.32 ─ 15
Use Multi Level Correction - User Correction
Activates the use of the multi level correction data (see chapter 5.5.9.3, "Filling the Cor-
rection List with Power Sensor Measurement Data", on page 230) if checked. Otherwise
the Single Level Correction measurement data is used.
SCPI command:
[:SOURce<hw>]:CORRection:MLEVel:STATe on page 707
User Correction Value - User Correction
Indicates the current value for level correction.
SCPI command:
[:SOURce<hw>]:CORRection:VALue on page 707
User Cor. Data - User Correction
Calls the "File Select" menu for selecting and creating a list or the "File Manager".
SCPI command:
[:SOURce]:CORRection:CSET:DELete on page 702
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 706
Edit User Cor. Data - User Correction
Calls the editor for editing the selected user correction list.
A list consists of any number of frequency/level value pairs. The currently selected list is
displayed.
Each list is saved as a separate file with extension *.uco. The file name and the directory
to which the file is saved are user-selectable.
Multi level lists which are filled automatically using the Fill with sensor function consist of
nine level columns, one for each attenuator stage (see chapter 5.5.9.3, "Filling the Cor-
rection List with Power Sensor Measurement Data", on page 230).
Note: Save list only after filling both columns (frequency and level), otherwise the entries
are lost.
RF Signal and Analog Modulations - RF Block