Instrument Function
R&S
®
SMU200A
188Operating Manual 1007.9845.32 ─ 15
Example:
50 bits were set to "0" by the DUT. These 50 bits plus the preceding "0" are ignored in
the bit error rate measurement.
"Off"
Pattern Ignore is not active.
"All 1"
Bit sequences consisting of 32 or more consecutive "1" data are not used
(i.e. ignored) for the BER measurement.
"All 0"
Bit sequences consisting of 32 or more consecutive "0" data are not used
(i.e. ignored) for the BER measurement.
SCPI command:
:BERT:SETup:IGNore on page 518:BERT:SETup:DENable on page 517
5.4.11 Block Error Rate Menu
The "Block Error Rate" menu for configuring the block error rate measurement is called
either in the "BERT" function block or via the MENU button under "BERT."
The top section of the "Block Error Rate" menu is used to switch on the block error rate
measurement and display the results and current status of the measurement.
The "Configuration" section is used to select the type of measurement and the termination
criteria and to set the polarity of the input signals.
Bit and Block Error Rate Measurements - BERT Block