EasyManua.ls Logo

Rohde & Schwarz SMU200A - Baseband Signal Level Measurement

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMU200A
319Operating Manual 1007.9845.32 ─ 15
Crest[dB]=Peak_Level[dBFs]/Signal_Level[dBFs]
SCPI command:
[:SOURce<hw>]:BB:CFACtor on page 689
Peak Level - Digital Impairments
Indicates the peak level of the baseband signal relative to full scale of 0.5 V (in terms of
dB full scale).
SCPI command:
[:SOURce<hw>]:BB:POWer:PEAK on page 690
Signal Level (RMS) - Digital Impairments
Indicates the rms level of the baseband signal relative to full scale of 0.5 V (in terms of
dB full scale).
SCPI command:
[:SOURce<hw>]:BB:POWer:RMS on page 690
5.7.4.2 Baseband Signal Level Measurement
Information on the baseband I/Q signal during measurement is indicated in the section
"Baseband Signal Level Measurement". Level, crest factor and duration of the measure-
ment can be displayed for the signal either with or without interfering.
State - Digital Impairments - Measurement
Activates/deactivates the measurement of the digital I/Q baseband signal.
When activated, information on the peak level, rms level and crest factor are indicated.
Additionally the duration of the measurement can be set.
SCPI command:
[:SOURce<hw>]:BB:MEASurement[:STATe] on page 688
Signal Source - Digital Impairments - Measurement
Selects the signal for the measurement. The signal level and crest factor can be retrieved
either from the interference-free digital I/Q baseband signal or the signal superimposed
by the noise generator AWGN.
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block

Table of Contents

Other manuals for Rohde & Schwarz SMU200A

Related product manuals