Remote Control Commands
R&S
®
SMU200A
700Operating Manual 1007.9845.32 ─ 15
CORR:CSET 'D:/Lists/ucor/ucor1'
selects the table ucor1.
CORR:CSET:DATA:FREQ 100MHz,102MHz,103MHz,...
enters the frequency value in the table UCOR1.
Example:
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
The command queries the number of frequency values in the selected table.
The numerical suffix at SOURce must not be used for this command.
Return values:
float<Points>
CORR:CSET 'D:/Lists/ucor/ucor1'
selects the table ucor1.
CORR:CSET:DATA:FREQ:POIN?
queries the number of frequency values in the table ucor1.
Response: 440
the table ucor1 contains 440 frequency values.
Example:
Query onlyUsage:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
The command transfers the level data to the table selected
with :CORRection:CSET:SELect.
*RST does not affect data lists. The numerical suffix at SOURce must not be used for
this command.
Parameters:
string
-40 dB to 6 dB
dB
<Power>
Range:
Default unit:
CORR:CSET 'D:/Lists/ucor/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW 1dB, 0.8dB, 0.75dB,...
enters the level values in the table ucor1.
Example:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
The command queries the number of level values in the selected table.
The numerical suffix at SOURce must not be used for this command.
Return values:
float<Points>
SOURce Subsystem