Remote Control Commands
R&S
®
SMU200A
798Operating Manual 1007.9845.32 ─ 15
:TEST<hw>:DIRect <HW_assembly>,<subadress>,<hex data string>
:TEST<hw>:DIRect? <HW_assembly>,<subadress>
The respective hardware assembly responds directly to the command; any safety mech-
anisms are bypassed. This function is only available via remote control.
Parameters for setting and query:
'SSYN' | 'IQOP3' | ' IQOP6' | ' SATT3' | 'SATT6' | 'DIRECTLAST'<HW_assembly>
TEST:DIR 'SSYN',0,#H12345678
TEST:DIR? 'SSYN',0
Response: #H12345678
Example:
:TEST:FADer?
This command performs a selftest on the fading simulator and the hardware options of
the fading simulator extension.
Return values:
0|1|OFF|ON<Fader>
TEST:FAD?
Response: "0" on success, "1" on fail
Example:
Query onlyUsage:
7.19 TRIGger Subsystem
The TRIGger system contains the commands for selecting the trigger source for the RF
and LF sweep. The trigger input connectors are configured in the SOURce:INPut sub-
system.
The trigger system of the R&S SMU is a simplified implementation of the SCPI trigger
system. The TRIGger system differs from the SCPI system as follows:
●
No INITiate command; the instrument behaves as if INITiate:CONTinuous ON
were set.
●
Under TRIGger several sweep subsystems exist.
Other commands associated with the trigger system of the R&S SMU can be found in
the modulation and RF signal subsystems.
The LF output is activated with TRIGger0.
TRIGger Subsystem