EasyManua.ls Logo

Rohde & Schwarz SMU200A - Selects an External Analog Signal as the Input Signal. the Signal Must be Applied at the Inputs I and Q

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
726Operating Manual 1007.9845.32 ─ 15
IQ:IMP OFF
deactivates I/Q impairment.
Example:
[:SOURce<hw>]:IQ:SOURce <Source>
This command selects the input signal for the I/Q modulator.
In case of two-path instruments, this selection is only available for Path A (SOURce[1]).
The input signal for the I/Q modulator of Path B is always the internal baseband signal
(BASeband).
Selecting ANALog (Wideband I/Q In) switches off:
SOURce:AM:STATe OFF
SOURce:BBAM:STATe OFF
SOURce:BB:DM:STATe OFF
SOURce:BB:<Dig. Standard>:STATe OFF
SOURce:BBIN:STATe OFF
Parameters:
BASeband|ANALog
BASeband
<Source>
*RST:
IQ:SOUR ANAL
selects an external analog signal as the input signal. The signal
must be applied at the inputs I and Q.
Example:
[:SOURce<hw>]:IQ:STATe <State>
This command activates and deactivates I/Q modulation.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
IQ:STAT ON
activates I/Q modulation.
Example:
[:SOURce<hw>]:IQ:SWAP[:STATe] <State>
When set to ON, this command swaps the I and Q channel.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
IQ:SWAP ON
swaps the I and Q channel.
Example:
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMU200A