Instrument Function
R&S
®
SMU200A
401Operating Manual 1007.9845.32 ─ 15
I Value (Const IQ Test Signal)
Enters the value for the I component of the test signal.
SCPI command:
[:SOURce<hw>]:BB:ARB:TSIGnal:CIQ:I on page 595
Q Value (Const IQ Test Signal)
Enters the value for the Q component of the test signal.
SCPI command:
[:SOURce<hw>]:BB:ARB:TSIGnal:CIQ:Q on page 595
Generate Signal HD
Generates a signal and saves it to a file.
The File Select window opens automatically and the signal can be stored as a waveform
file.
SCPI command:
[:SOURce<hw>]:BB:ARB:TSIGnal:CIQ:CREate:NAMed on page 595
Generate Signal RAM
Generates a signal and uses it as output straight away.
SCPI command:
[:SOURce<hw>]:BB:ARB:TSIGnal:CIQ:CREate on page 594
5.8.4.2 Multi Segment Waveforms
To fulfill the requirements of some test systems and to enable rapid alternation between
different waveforms with differing test signals, the R&S Signal Generator provides the
functionality to generate multi segment waveform files. A multi segment waveform is a
combination of multiple waveforms called segments. Each segment represents a com-
pletely independent waveform that can be output with its own marker and clock settings.
The entire multi segment waveform is loaded into memory. The segment intended to be
output at any given moment can be selected by the user. It is therefore possible to alter-
nate between the individual waveform segments without experiencing any delay due to
the loading operation.
If very high switchover speeds are required, the test signals can be continuously scrolled
through with the aid of an external trigger or by applying a predefined “play list”. For this
purpose the segments must have a common sample rate. If the combined waveforms
have different sample rates, they can be adapted to a common sample rate by resam-
pling. By the same procedure the instantaneous amplitude of the various waveforms can
be scaled to a common RMS level.