Remote Control Commands
R&S
®
SMU200A
564Operating Manual 1007.9845.32 ─ 15
SENSe<ch>[:POWer]:SOURce <Source>
The command selects the signal source for the measurement.
In case of two-path instruments, the RF signal of path A or path B or a user defined source
can be selected.
Parameters:
A|B|USER
A
<Source>
*RST:
SENS:SOUR A
selects the RF signal as measurement source. The RF frequency
is used as the measurement frequency of the sensor and the cor-
responding correction factor is used. The level setting of the instru-
ment serves as reference level of the measurement.
Example:
SENSe<ch>[:POWer]:STATus[:DEVice]?
The command queries if a sensor is connected to the signal generator.
The sensor is selected by suffix in the keyword SENSe or READ of the command header.
Suffix 1 denotes the sensor connected to the SENSOR connector, suffix 2 the sensor
connected first to one of the USB interfaces and suffix 3 the sensor connected second
to one of the USB interfaces.
Return values:
0|1|OFF|ON<DEVice>
SENS:STAT?
queries if a sensor is connected to the instrument.
Response: 1
a sensor is connected to the POWER SENSOR interface.
Example:
Query onlyUsage:
SENSe<ch>[:POWer]:SVERsion?
The command queries the software version of the connected R&S NRP power sensor.
Return values:
string<Sversion>
SENS:POW:SVER?
queries the software version of the R&S NRP power sensor.
Example:
Query onlyUsage:
SENSe, READ and INITiate Subsystems