Remote Control Commands
R&S
®
SMU200A
633Operating Manual 1007.9845.32 ─ 15
Parameters:
AUTO|RETRigger|AAUTo|ARETrigger|SINGle
AUTO
The modulation signal is generated continuously.
RETRigger
The modulation signal is generated continuously. A trigger event
(internal or external) causes a restart.
AAUTO
The modulation signal is generated only when a trigger event
occurs. After the trigger event the signal is generated
contunuously. Signal generation is stopped with comman
SOUR:BB:DM:TRIG:ARM:EXEC and started again when a trigger
event occurs.
ARETrigger
The modulation signal is generated only when a trigger event
occurs. THe device automatically toggles to RETRIG mode. Every
subsequent trigger event causes a restart. Signal generation is
stopped with comman SOUR:BB:DM:TRIG:ARM:EXEC and
started again when a trigger event occurs.
SINGle
The modulation signal is generated only when a trigger event
occurs. After the trigger event the signal is generated once to the
set sequence length (SOUR:BB:DM:TRIG:SLEN). Every
subsequent trigger event caused a restart.
AUTO
<Sequence>
*RST:
BB:DM:SEQ AAUT
sets the "Armed_auto" trigger mode; the device waits for the first
trigger (e.g. with *TRG) and then generates the signal continu-
ously.
Example:
Marker Settings
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE......................................................633
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:OFFTime..................................................634
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:ONTime....................................................635
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PATTern...................................................635
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:DIVider..........................................635
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:FREQuency...................................636
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE <Mode>
The command defines the signal for the selected marker output.
SOURce Subsystem