EasyManua.ls Logo

Rohde & Schwarz SMU200A - Marker Settings

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
633Operating Manual 1007.9845.32 ─ 15
Parameters:
AUTO|RETRigger|AAUTo|ARETrigger|SINGle
AUTO
The modulation signal is generated continuously.
RETRigger
The modulation signal is generated continuously. A trigger event
(internal or external) causes a restart.
AAUTO
The modulation signal is generated only when a trigger event
occurs. After the trigger event the signal is generated
contunuously. Signal generation is stopped with comman
SOUR:BB:DM:TRIG:ARM:EXEC and started again when a trigger
event occurs.
ARETrigger
The modulation signal is generated only when a trigger event
occurs. THe device automatically toggles to RETRIG mode. Every
subsequent trigger event causes a restart. Signal generation is
stopped with comman SOUR:BB:DM:TRIG:ARM:EXEC and
started again when a trigger event occurs.
SINGle
The modulation signal is generated only when a trigger event
occurs. After the trigger event the signal is generated once to the
set sequence length (SOUR:BB:DM:TRIG:SLEN). Every
subsequent trigger event caused a restart.
AUTO
<Sequence>
*RST:
BB:DM:SEQ AAUT
sets the "Armed_auto" trigger mode; the device waits for the first
trigger (e.g. with *TRG) and then generates the signal continu-
ously.
Example:
Marker Settings
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE......................................................633
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:OFFTime..................................................634
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:ONTime....................................................635
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PATTern...................................................635
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:DIVider..........................................635
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:FREQuency...................................636
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE <Mode>
The command defines the signal for the selected marker output.
SOURce Subsystem