Remote Control Commands
R&S
®
SMU200A
492Operating Manual 1007.9845.32 ─ 15
*PRE <Value>
Parallel poll Register Enable
Sets parallel poll enable register to the indicated value. The query returns the contents
of the parallel poll enable register in decimal form.
Parameters:
0 to 255
<Value>
Range:
*PSC <Action>
Power on Status Clear
Determines whether the contents of the ENABle registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
<Action>
*RCL <Number>
RECALL calls up the instrument status which was stored under the specified number
using the *SAV command, e.g. *SAV 4. It also activates the instrument settings which
are stored in a file and loaded using the MMEMory:LOAD <number>,
<file_name.extension> command.
*RST
ReSeT
Sets the instrument to a defined default status. It is equivalent to SYSTem:PRESet. The
default settings are indicated in the description of commands.
Setting onlyUsage:
Common Commands