Remote Control Commands
R&S
®
SMU200A
723Operating Manual 1007.9845.32 ─ 15
[:SOURce]:INPut:TRIGger:SLOPe <Slope>
Sets the polarity of the active slope of an applied instrument trigger.
This setting affects the INST TRIG input (BNC connector at the rear of the instrument).
For two-path instruments, the external trigger for path B is applied at the INST TRIG pin
at the AUX I/O interface.
Parameters:
NEGative|POSitive
POSitive
<Slope>
*RST:
INP:TRIG:SLOP NEG
the active slope of the external trigger signal at the INST TRIG
input is the falling slope.
Example:
7.15.9
SOURce:IQ Subsystem
This subsystem contains the commands for checking the I/Q modulation.
[:SOURce<hw>]:IQ:CREStfactor......................................................................................723
[:SOURce<hw>]:IQ:GAIN................................................................................................724
[:SOURce<hw>]:IQ:IMPairment:IQRatio[:MAGNitude]........................................................724
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I.......................................................................724
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q.....................................................................725
[:SOURce<hw>]:IQ:IMPairment:QUADrature:[ANGLe].......................................................725
[:SOURce<hw>]:IQ:IMPairment[:STATe]..........................................................................725
[:SOURce<hw>]:IQ:SOURce...........................................................................................726
[:SOURce<hw>]:IQ:STATe..............................................................................................726
[:SOURce<hw>]:IQ:SWAP[:STATe].................................................................................726
[:SOURce<hw>]:IQ:WBSTate..........................................................................................727
[:SOURce<hw>]:BB:IQGain............................................................................................727
[:SOURce<hw>]:IQ:CREStfactor <Crestfactor>
This command specifies the crest factor of the external analog signal.
Parameters:
float
0dB to 30dB
0.01dB
0dB
dB
<Crestfactor>
Range:
Increment:
*RST:
Default unit:
IQ:CRES 10
specifies a crest factor of 10 dB for the external analog signal.
Example: