Remote Control Basics
R&S
®
SMU200A
481Operating Manual 1007.9845.32 ─ 15
All status registers have the same internal structure.
SRE, ESE
The service request enable register SRE can be used as ENABle part of the STB if the
STB is structured according to SCPI. By analogy, the ESE can be used as the ENABle
part of the ESR.
6.5.2 Structure of a SCPI Status Register
Each standard SCPI register consists of 5 parts. Each part has a width of 16 bits and has
different functions. The individual bits are independent of each other, i.e. each hardware
status is assigned a bit number which is valid for all five parts. Bit 15 (the most significant
bit) is set to zero for all parts. Thus the contents of the register parts can be processed
by the controller as positive integers.
Fig. 6-2: The status-register model
Description of the five status register parts
The five parts of a SCPI register have different properties and functions:
●
CONDition
The CONDition part is written into directly by the hardware or the sum bit of the next
lower register. Its contents reflect the current instrument status. This register part can
only be read, but not written into or cleared. Its contents are not affected by reading.
●
PTRansition
The two transition register parts define which state transition of the CONDition part
(none, 0 to 1, 1 to 0 or both) is stored in the EVENt part.
The Positive-TRansition part acts as a transition filter. When a bit of the
CONDition part is changed from 0 to 1, the associated PTR bit decides whether the
EVENt bit is set to 1.
- PTR bit =1: the EVENt bit is set.
- PTR bit =0: the EVENt bit is not set.
Status Reporting System