Remote Control Commands
R&S
®
SMU200A
790Operating Manual 1007.9845.32 ─ 15
Parameters:
float
1 to 30
28
<Address>
Range:
*RST:
SYST:COMM:GPIB:ADDR 14
sets GPIB address.
Example:
SCPI conformUsage:
:SYSTem:COMMunicate:NETWork[:COMMon]:DOMain <Domain>
The command enters the domain of the network.
Parameters:
string<Domain>
SYST:COMM:NETW:COMM:DOM ABC.DE
enters the domain of the network.
Example:
:SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup <Workgroup>
The command enters the individual workgroup name of the instrument.
Parameters:
string<Workgroup>
SYST:COMM:NETW:COMM:WORK 'TEST_09'
enters the workgroup name 'TEST_09'.
Example:
:SYSTem:COMMunicate:NETWork:RESource? <Resource>
The command queries the visa resource string. This string is used for remote control of
the instrument.
Parameters:
string<Resource>
SYST:COMM:NETW:RES
queries the VISA resource string.
Response: TCPIP::192.1.2.3::INSTR
Example:
Query onlyUsage:
:SYSTem:COMMunicate:GPIB:RESource?
The command queries the visa resource string for remote control via the GPIB interface.
This string is used for remote control of the instrument.
To change the GPIB address, use the command SYST:COMM:GPIB:ADDR.
SYSTem Subsystem