Remote Control Commands
R&S
®
SMU200A
793Operating Manual 1007.9845.32 ─ 15
SYSTem:ERRor:COUNt?
The command queries the number of entries in the error queue. If the error queue is
empty, '0' is returned.
Return values:
string<Count>
SYST:ERR:COUN
queries the number of entries in the error queue.
Response: 1
One error has occurred since the error queue was last read out.
Example:
Query only
SCPI conform
Usage:
SYSTem:ERRor[:NEXT]?
The command queries the oldest entry in the error queue and then deletes it.
The command is identical to the command STATus:QUEue:NEXT.
Return values:
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by SCPI
(see section chapter 9, "Error Messages", on page 829)
<Next>
SYST:ERR?
queries the oldest entry in the error queue.
Response: 0, 'no error'
No errors have occurred since the error queue was last read out.
Example:
Query only
SCPI conform
Usage:
:SYSTem:KLOCk <Klock>
This command (Keyboard LOCk) disables the front panel keyboard of the instrument
including the LOCAL key, or enables it again (OFF).
Parameters:
0|1|OFF|ON
OFF
<Klock>
*RST:
SYSTem Subsystem