Instrument Function
R&S
®
SMU200A
218Operating Manual 1007.9845.32 ─ 15
the menu. The attenuator mode is set in the "Attenuator Settings" section. The step width
which is used when setting the level using the rotary knob (with "Variation Active On") is
set in the "User Variation" section.
The power-on behavior of the instrument and the level display in the display header are
set in the "Power-On / EMF Settings" section (see chapter 5.5.7.3, "Power-On/EMF Set-
tings", on page 220).
The lower sections provide access to the "Automatic Level Control" settings and to func-
tion User Correction settings. For more information about this functions, see chap-
ter 5.5.8, "Automatic Level Control - ALC", on page 221 and chapter 5.5.9, "User Cor-
rection", on page 223.
Level Settings
The offset-free level, attenuation mode, level offset and level limit are set in the top section
of the menu.
Amplitude - RF Signal
Sets the RF level of the RF output connector.
The level entered and displayed here corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note: The SCPI command SOUR:POW:LEV:IMM:AMPL sets the level of the "Level" dis-
play, i.e. the level containing offset.
SCPI command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude] on page 765
Limit - RF Signal
Sets the level limit.
The value specifies the upper limit of the level at the RF output connector. A message
appears if an attempt is made to set a level above this limit and the level at the RF output
is confined to the upper limit. However, the level indication is not influenced.
RF Signal and Analog Modulations - RF Block