Instrument Function
R&S
®
SMU200A
310Operating Manual 1007.9845.32 ─ 15
Optimize IQ-Signals for RF Output
Activates/deactivates optimization of level accuracy and imbalance for RF Output.
Level accuracy and imbalance can be either optimized for RF output (active) or for dif-
ferential outputs (not active).
With a second option Baseband Main Module (R&S SMU-B13) it is possible to optimize
a signal for RF output in one path and for the differential output in the other path.
SCPI command:
[:SOURce<hw>]:IQ:OUTPut:BIAS:OPTimization[:RF]:STATe on page 730
I/Q Output Type
Selects the type of output. The analog outputs can be operated in "Single Ended" mode
or "Differential" (non-inverted) mode. The menu changes depending on the selected out-
put mode
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block