EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 515

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
495Operating Manual 1007.9845.32 ─ 15
SOUR:PRES
presets all settings that are related to signal path
Example:
EventUsage:
SYSTem:PRESet <Preset>
The command triggers an instrument reset. It has the same effect as the PRESET key
on the front panel and the *RST command.
For an overview of the settings affected by the preset function, see chapter 5.2.2, "Default
Instrument Settings - Preset Key", on page 126.
Return values:
string<Preset>
SYST:PRES
All instrument settings (also those that are not currently active) are
reset to their default values.
Example:
Event
SCPI conform
Usage:
:SYSTem:FPReset
The command triggers an instrument reset to the original state of delivery.
Note: Since Factory Preset resets the Remote Channel and network settings to the
default values, executing Factory Preset via remote control terminates the connection to
the instrument, if these settings had been configured to values different to the default
ones!
The factory preset function resets nearly all instrument settings. In addition to the regular
preset by means of the PRESET key, a "Factory Preset" resets also the following values:
Reference frequency settings ("Ref Oscillator" menu)
Power on settings ("Level/EMF" menu)
Network settings including hostname ("Setup" menu)
Remote Channel settings including GPIB address ("Setup" menu)
Start/Stop Gui Update "(Setup" menu)
Display and keyboard settings ("Setup" menu).
To maintain security, password settings and all settings protected by these passwords
like disabled USB and LAN connections are not changed.
Not affected by the "Factory Preset" are also user data, lists or instrument settings files,
created for example by means of the Save/Recall function.
SYST:FPR
all instrument settings (also those that are not currently active) are
reset to the factory values.