{MARKER LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'MARKER LIST 1'
queries the content of the MARKER LIST 1 tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
the marker setting for samples 0 to 9 = 0 (low), for 10 to 19 = 1
(high) and for 20 to 29 = 0. From sample 30 onward the marker
setting = 1.
{LEVATT LIST 1: 0:0;10:1;20:0;30:1}
BB:DM:CLIS:TAG? 'LEVATT LIST 1'
queries the content of the LEVATT LIST 1 tag of the selected
control list file.
Response: '0:0;10:1;20:0;30:1'
level attenuation applies to data values 10 to 19 (high) and from
data value 30 onward.