EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 600

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
580Operating Manual 1007.9845.32 ─ 15
Parameters:
float
0.1 dB
0 dB
<Pgain>
Increment:
*RST:
BB:PGA 3dB
sets a relative path gain of 3 dB for the baseband signal of path A
compared to the baseband signal of path B.
Example:
[:SOURce<hw>]:BB:POFFset <Poffset>
The command sets the relative phase offset for the selected baseband signal compared
to the baseband signals of the other baseband sources (second path or external base-
band).
The phase offset affects the signal on the "baseband block" output.
Parameters:
float
0 to 359.9
0.1 DEG
0 DEG
<Poffset>
Range:
Increment:
*RST:
BB:POFF 0.5DEG
sets a relative phase offset of 0.5 DEG for the baseband signal
Example:
[:SOURce<hw>]:BB:ROUTe <Route>
The command selects the signal route for the baseband signal of a two-path instrument.
Parameters:
A|B|AB
A
The baseband signal is introduced into path A. The signals from
both paths are summed if necessary.
B
The baseband signal is introduced into path B. The signals from
both paths are summed if necessary.
AB
The baseband signal is introduced into path A and path B. The
signals from both paths are summed if necessary.
A
<Route>
*RST:
SOUR2:BB:ROUT A
the baseband signal of path B is introduced into path A.
Example:
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMU200A

Related product manuals