Remote Control Commands
R&S
®
SMU200A
600Operating Manual 1007.9845.32 ─ 15
BB:ARB:MCAR:CARR15:FILE 'D:/IQ_wcdma'
selects file IQ_wcdma. The data of the file is modulated onto car-
rier 15.
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:PHASe <Phase>
The command sets the start phase of the selected carrier.
The phase settings are only valid if optimization of the crest factor is disabled
(:SOURce:BB:ARB:MCARrier:CFACtor:MODE OFF).
Parameters:
float
0.0 to 360.0
0
DEG
<Phase>
Range:
*RST:
Default unit:
BB:ARB:MCAR:CARR15:PHAS 90 DEG
sets a start phase of 90° for carrier 15.
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:POWer <Power>
The command sets the gain of the selected carrier.
Parameters:
float
-80.0 to 0.0
0.01
0
dB
<Power>
Range:
Increment:
*RST:
Default unit:
BB:ARB:MCAR:CARR15:POW -50 dB
sets the power of carrier 15 to -50 dB.
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:STATe <State>
The command switches the selected carrier on or off.
Parameters:
0|1|OFF|ON
ON
<State>
*RST:
BB:ARB:MCAR:CARR15:STAT ON
switches carrier 15 on.
Example:
SOURce Subsystem