Remote Control Commands
R&S
®
SMU200A
627Operating Manual 1007.9845.32 ─ 15
BB:DM:TRIG:SOUR INT
sets internal triggering.
BB:DM:TRIG:SEQ ARET
sets Armed_Retrigger mode, i.e. every trigger event causes signal
generation to restart.
BB:DM:TRIG:EXEC
executes a trigger, signal generation is started.
BB:DM:TRIG:ARM:EXEC
signal generation is stopped.
BB:DM:TRIG:EXEC
executes a trigger, signal generation is started again.
Example:
EventUsage:
[:SOURce<hw>]:BB:DM:TRIGger:EXECute
The command executes a trigger. The internal trigger source must be selected using the
command:BB:DM:TRIG:SOUR INT and a trigger mode other than AUTO must be
selected using the command :BB:DM:TRIG:SEQ.
BB:DM:TRIG:SOUR INT
sets internal triggering.
BB:DM:TRIG:SEQ RETR
sets Retrigger mode, i.e. every trigger event causes signal gen-
eration to restart.
BB:DM:TRIG:EXEC
executes a trigger.
Example:
EventUsage:
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
(enabled for "Trigger Source" External)
Enables/disables output of the signal synchronous to the external trigger event.