EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 712

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
692Operating Manual 1007.9845.32 ─ 15
[:SOURce]:BBIN:FOFFset <Foffset>
This command enters the frequency offset for the external baseband signal. The offset
affects the signal on the output of the "Baseband" block.
The complex I/Q bandwidth of the shifted useful signal must not exceed 80 MHz in total.
The following applies:
f
offset
-(f
use
/2) => -(80/2) MHz and f
offset
+(f
use
/2) >= +(80/2) MHz
f
use
= the complex useful bandwidth of the I/Q signal before the offset.
f
offset
= frequency offset.
Parameters:
float
-40 to +40
0.01Hz
0Hz
Hz
<Foffset>
Range:
Increment:
*RST:
Default unit:
BBIN:FOFF 2 MHz
sets a frequency offset of 2 MHz.
Example:
[:SOURce<hw>]:BBIN:GIMBalance <Gimbalance>
This command enters a gain to the Q component of the external baseband signal.
Parameters:
float
-3.0dB to +3.0dB
0.01dB
0dB
dB
<Gimbalance>
Range:
Increment:
*RST:
Default unit:
BBIN:GIMB -1.0dB
amplifies the Q component of the external input signal with -1.0
dB.
Example:
[:SOURce]:BBIN:IQSWap[:STATe] <State>
This command swaps the I and Q channel if set to ON.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
BBIN:IQSW ON
swaps the I and Q channel of the external baseband signal.
Example:
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMU200A

Related product manuals