Remote Control Commands
R&S
®
SMU200A
718Operating Manual 1007.9845.32 ─ 15
Parameters:
float
0 to RFmax
0.01Hz
400MHz
<Span>
Range:
Increment:
*RST:
FREQ:CENT 400 MHz
sets the center frequency of the frequency sweep to 400 MHz.
FREQ:SPAN 200 MHz
sets a span of 200 MHz. This sets the sweep range to 300 MHz
to 500 MHz.
Example:
SCPI conformUsage:
[:SOURce<hw>]:FREQuency:STARt <Start>
This command sets the start frequency for the sweep mode. STARt can be greater than
STOP.
This command is linked to the commands [SOUR]:FREQ:CENT and
[SOUR]:FREQ:SPAN, i.e. changing these values causes the STARt value to change,
and vice versa:
STARt = (CENTer - SPAN/2).
As with the Frequency value entered in the header, the OFFSet value is also taken into
consideration with this command. The specified value range is therefore only effective if
OFFSet is set to 0. The value range for other OFFset values can be calculated using
the following formula:
300 kHz + OFFSet ... RF
max
+ OFFSet
Correlation: STARt = (CENTer - SPAN/2)
Parameters:
float
300kHz to RFmax
0.01Hz
100MHz
<Start>
Range:
Increment:
*RST:
FREQ:STARt 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
Example:
SCPI conformUsage:
SOURce Subsystem