Remote Control Commands
R&S
®
SMU200A
771Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:POWer:STEP:MODE <Mode>
This command activates (USER) or deactivates (DECimal) the user-defined step width
used when varying the level value with the level values UP/DOWN. The command is linked
to setting "Variation Active" for manual control, i.e. the command also activates/deacti-
vates the user-defined step width used when varying the level value with the rotary knob.
In the case of two-path instruments, the step width can be set separately for each path.
Parameters:
DECimal|USER
DECimal
<Mode>
*RST:
POW:STEP 2
sets the step width for the level setting to 2 dB.
POW:STEP:MODE USER
actives this step width for level variation with the rotary knob (man-
ual control) and with level values UP/DOWN (remote control).
Example:
[:SOURce<hw>]:POWer:STOP <Stop>
The command sets the stop level in Sweep mode.
As with the Level value entered in the RF Level menu, the OFFSet value is also taken
into consideration with this command.
The specified value range is therefore only effective if:SOURce:POWer:OFFSet is set to
0. The value range for other OFFset values can be calculated using the following formula:
Minimum level + OFFSet ...Maximum level + OFFSet
Parameters:
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The value range depends on whether or not the instrument is
equipped with the option High Power.
The values are given in the data sheet.
0.01 dB
-10 dBm
dBm
<Stop>
Increment:
*RST:
Default unit:
POW:STOP 3
sets the stop level for the level sweep to 3 dBm for RF output A.
Example:
SCPI conformUsage:
SOURce Subsystem