EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #152 background imageLoading...
Page #152 background image
Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 2:
General operation
DMM7510-901-01 Rev. B / May 2015 2-125
Another example is a 1 kΩ/1 kΩ/1 kΩ resistor network. If 3 V is applied across the total three 1 kΩ
resistors and V
sense
is across the first 1 kΩ resistor, set V
input
to the 10 V range and V
sense
to the 1 V
range. The ratio measurement is approximately 3.00000. If V
sense
is set to the 1 V range, the ratio
displays overflow, with SENSE HI and SENSE LO terminals exceeding the 125 % maximum
reference to the LO terminals. The SENSE HI to LO is 3 V and SENSE LO to LO is 2 V, respectively.
Figure 89: DCV ratio 1 kohm/1 kohm/1 kohm resistor network
Do not apply more than 1000 V
peak
to the INPUT terminals or more than 350 V
peak
to the SENSE
terminals. Failure to heed this caution may result in instrument damage.
DC voltage ratio measure connections
Figure 90: Front panel connections: DC voltage ratio measurement

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals