2: General operation Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
2-126 DMM7510-901-01 Rev. B / May 2015
Figure 91: Rear panel connections: DC voltage ratio measurement
Measure DC voltage ratio using the front panel
To make a DC voltage ratio measurement using the front panel:
1. Make the connections as shown in the previous figures.
2. Press the FUNCTION key.
3. Select DCV Ratio.
4. Press the TRIGGER key for 2 seconds and verify that the instrument is set to Continuous
Measurement.
The measurements start displaying on the front panel.
Settings available for DC voltage ratio measurements
See DC voltage ratio measure settings (on page 2-32) for settings that are available when you are
making continuity measurements.