EasyManuals Logo

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #296 background imageLoading...
Page #296 background image
Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 3:
Functions and features
DMM7510-901-01 Rev. B / May 2015 3-113
Using the data queue for real-time communication
Nodes that are running test scripts at the same time can store data in the data queue for real-time
communication. Each instrument has an internal data queue that uses the first-in, first-out (FIFO)
structure to store data. You can use the data queue to post numeric values, strings, and tables.
Use the data queue commands to:
Share data between test scripts running in parallel
Access data from a remote group or a local node on a TSP-Link
®
network at any time
You cannot access the reading buffers or global variables from any node in a remote group while a
node in that group is performing an overlapped operation. However, you can use the data queue to
retrieve data from any node in a group that is performing an overlapped operation. In addition, the
master node and the group leaders can use the data queue as a way to coordinate activities.
Tables in the data queue consume one entry. When a node stores a table in the data queue, a copy
of the data in the table is made. When the data is retrieved from the data queue, a new table is
created on the node that is retrieving the data. The new table contains a completely separate copy of
the data in the original table, with no references to the original table or any subtables.
You can access data from the data queue even if a remote group or a node has overlapped
operations in process. See the dataqueue commands in the TSP command reference (on page 8-1
)
for more information.
Remote TSP-Link commands
Commands that control and access the TSP-Link
®
synchronization port are summarized in the
following table. See the TSP command reference (on page 8-1
) for complete details on these
commands.
Use the commands in the following table to perform basic steady-state digital I/O operations; for
example, you can program the Model DMM7510 to read and write to a specific TSP-Link
synchronization line or to the entire port.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals