6: SCPI command reference Model DMM7510 Digit Graphical Sampling Multimeter
Reference Manual
6-100 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:RTD:FOUR
This command contains the type of 4-wire RTD that is being used.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
PT100
Usage
[:SENSe[1]]:<function>:RTD:FOUR <type>
[:SENSe[1]]:<function>:RTD:FOUR?
<function>
The function to which the setting applies; see Functions
<type>
The type of four-wire RTD:
PT100: PT100
PT385: PT385
PT3916: PT3916
D100: D100
F100: F100
User-specified type: USER
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
The transducer type must be set to temperature and the transducer must be set to 4-wire RTD before
you can set the RTD type.
Example
:FUNC "TEMP"
:TEMP:TRANsducer FRTD
:TEMP:RTD:FOUR PT3916
Set the measure function to temperature.
Set the transducer type to 4-wire RTD.
Set the RTD type to PT3916.
Also see
[:SENSe[1]]:<function>:TRANsducer (on page 6-112)