Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6: SCP
I command reference
DMM7510-901-01 Rev. B / May 2015 6-101
[:SENSe[1]]:<function>:RTD:THRee
This command defines the type of three-wire RTD that is being used.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
PT100
Usage
[:SENSe[1]]:<function>:RTD:THRee <type>
[:SENSe[1]]:<function>:RTD:THRee?
<function>
The function to which the setting applies; see Functions
<type>
The type of three-wire RTD:
• PT100: PT100
• PT385: PT385
• PT3916: PT3916
• D100: D100
• F100: F100
•
User-specified type: USER
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
The transducer type must be set to temperature and the transducer must be set to 3-wire RTD before
you can set the RTD type.
Example
:FUNC "TEMP"
:TEMP:TRANsducer TRTD
:TEMP:RTD:THR PT3916
Set the measure function to temperature.
Set the transducer type to 3-wire RTD.
Set the RTD type to PT3916.
Also see
[:SENSe[1]]:<function>:TRANsducer (on page 6-112)
Temperature measurements (on page 2-120)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals