6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter Refer
ence Manual
6-164 DMM7510-901-01 Rev. B / May 2015
Example
TRACe:MAKE "MyBuffer", 100
SENSe:COUNt 5
MEASure:CURRent:DC? "MyBuffer"
TRACe:DATA? 1,5, "MyBuffer", READ, REL
TRACe:SAVE "/usb1/myData.csv", "MyBuffer"
TRACe:SAVE "/usb1/myDataRel.csv", "MyBuffer", REL
Create a buffer called MyBuffer with a
maximum size of 100.
Make five readings for each
measurement request and return the
data.
Make the measurements.
Read the reading and relative timestamp
value for each point from 1 to 5.
Output:
-0.000000,0.000000,
-0.000000,
0.301759,-0.000000,0.579068,-
0.000000,
0.884302,-
0.000000,1.157444Save all reading
and default time information from a buffer
named MyBuffer to a file named
myData.csv on the USB flash drive.
Save all readings and relative
timestamps from MyBuffer to a file
named myDataRel.csv on the USB
flash drive.
Also see
Reading buffers (on page 3-13)
Remote buffer operation (on page 3-30)
:TRACe:MAKE (on page 6-160)
:TRACe:SAVE:APPend (on page 6-165)