Common (*) Commands 5
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 113
Query Syntax *ESE?
The *ESE? query returns the current contents of the Standard Event Status Enable
Register.
Return Format
<mask_argument><NL>
<mask_argument> ::= 0,..,255; an integer in NR1 format.
See Also "Introduction to Common (*) Commands" on page 109
"*ESR (Standard Event Status Register)" on page 114
"*OPC (Operation Complete)" on page 118
"*CLS (Clear Status)" on page 111
Table 40 Standard Event Status Enable (ESE)
Bit Name Description When Set (1 = High = True), Enables:
7 PON Power On Event when an OFF to ON transition occurs.
6 URQ User Request Event when a front-panel key is pressed.
5 CME Command Error Event when a command error is detected.
4 EXE Execution Error Event when an execution error is detected.
3 DDE Device Dependent Error Event when a device-dependent error is detected.
2 QYE Query Error Event when a query error is detected.
1 RQL Request Control Event when the device is requesting control. (Not
used.)
0 OPC Operation Complete Event when an operation is complete.