:MTESt Commands 21
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 405
:MTESt:AMASk:CREate
(see page 798)
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask around the
current selected channel, using the tolerance parameters defined by the
:MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and :MTESt:AMASk:UNITs
commands. The mask only encompasses the portion of the waveform visible on
the display, so you must ensure that the waveform is acquired and displayed
consistently to obtain repeatable results.
The :MTESt:SOURce command selects the channel and should be set before using
this command.
See Also • "Introduction to :MTESt Commands" on page 401
• ":MTESt:AMASk:XDELta" on page 408
• ":MTESt:AMASk:YDELta" on page 409
• ":MTESt:AMASk:UNITs" on page 407
• ":MTESt:AMASk:SOURce" on page 406
• ":MTESt:SOURce" on page 431
Example Code • "Example Code" on page 401

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals