754 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
31 Obsolete and Discontinued Commands
:SAVE:IMAGe:AREA
(see page 798)
Query Syntax
:SAVE:IMAGe:AREA?
The :SAVE:IMAGe:AREA? query returns the selected image area.
When saving images, this query returns SCR (screen). When saving setups or
waveform data, this query returns GRAT (graticule) even though graticule images
are not saved.
Return Format
<area><NL>
<area> ::= {GRAT | SCR}
See Also "Introduction to :SAVE Commands" on page 442
":SAVE:IMAGe[:STARt]" on page 445
":SAVE:IMAGe:FACTors" on page 446
":SAVE:IMAGe:FORMat" on page 447
":SAVE:IMAGe:INKSaver" on page 448
":SAVE:IMAGe:PALette" on page 449