286 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
16 :FRANalysis Commands
:FRANalysis:WGEN:LOAD
(see page 798)
Command Syntax
:FRANalysis:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :FRANalysis:WGEN:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
Query Syntax
:FRANalysis:WGEN:LOAD?
The :FRANalysis:WGEN:LOAD? query returns the current expected output load
impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also • ":FRANalysis:DATA" on page 275
• ":FRANalysis:ENABle" on page 276
• ":FRANalysis:FREQuency:STARt" on page 279
• ":FRANalysis:FREQuency:STOP" on page 280
• ":FRANalysis:RUN" on page 281
• ":FRANalysis:SOURce:INPut" on page 282
• ":FRANalysis:SOURce:OUTPut" on page 283
• ":FRANalysis:WGEN:VOLTage" on page 287

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals