Common (*) Commands 5
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 121
*RST (Reset)
(see page 798)
Command Syntax
*RST
The *RST command places the instrument in a known state. This is the same as
pressing [Save/Recall] > Default/Erase > Factory Default on the front panel.
When you perform a factory default setup, there are no user settings that remain
unchanged. To perform the equivalent of the front panel's [Default Setup] key,
where some user settings (like preferences) remain unchanged, use the
:SYSTem:PRESet command.
Reset conditions are:
Acquire Menu
Mode Normal
Averaging Off
# Averages 8
Analog Channel Menu
Channel 1 On
Channel 2 Off
Volts/division 5.00 V
Offset 0.00
Coupling DC
Probe attenuation 10:1
Vernier Off
Invert Off
BW limit Off
Impedance 1 M Ohm (cannot be changed)
Units Volts
Skew 0
Cursor Menu
Source Channel 1