:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 527
:SBUS<n>:UART:BITorder
(see page 798)
Command Syntax
:SBUS<n>:UART:BITorder <bitorder>
<bitorder> ::= {LSBFirst | MSBFirst}
The :SBUS<n>:UART:BITorder command specifies the order of transmission used
by the physical Tx and Rx input signals for the serial decoder and/or trigger when
in UART mode. LSBFirst sets the least significant bit of each message "byte" as
transmitted first. MSBFirst sets the most significant bit as transmitted first.
Query Syntax
:SBUS<n>:UART:BITorder?
The :SBUS<n>:UART:BITorder? query returns the current UART bit order setting.
Return Format
<bitorder><NL>
<bitorder> ::= {LSBF | MSBF}
See Also "Introduction to :TRIGger Commands" on page 585
":TRIGger:MODE" on page 594
":SBUS<n>:UART:TRIGger:TYPE" on page 542
":SBUS<n>:UART:SOURce:RX" on page 535
":SBUS<n>:UART:SOURce:TX" on page 536