660 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
28 :WAVeform Commands
:WAVeform:UNSigned
(see page 798)
Command Syntax
:WAVeform:UNSigned <unsigned>
<unsigned> ::= {{0 | OFF} | {1 | ON}}
The :WAVeform:UNSigned command turns unsigned mode on or off for the
currently selected waveform. Use the WAVeform:UNSigned command to control
whether data values are sent as unsigned or signed integers. This command can
be used to match the instrument's internal data type to the data type used by the
programming language. This command has no effect if the data format is ASCii.
If :WAVeform:SOURce is set to POD1, POD2, BUS1, or BUS2,
WAVeform:UNSigned must be set to ON.
Query Syntax
:WAVeform:UNSigned?
The :WAVeform:UNSigned? query returns the status of unsigned mode for the
currently selected waveform.
Return Format
<unsigned><NL>
<unsigned> ::= {0 | 1}
See Also • "Introduction to :WAVeform Commands" on page 635
• ":WAVeform:SOURce" on page 654

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals