EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:CHANnel<n> Commands 10
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 211
:CHANnel<n>:OFFSet
(see page 798)
Command Syntax
:CHANnel<n>:OFFSet <offset> [<suffix>]
<offset> ::= Vertical offset value in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:OFFSet command sets the value that is represented at center
screen for the selected channel. The range of legal values varies with the value set
by the :CHANnel<n>:RANGe and :CHANnel<n>:SCALe commands. If you set the
offset to a value outside of the legal range, the offset value is automatically set to
the nearest legal value. Legal values are affected by the probe attenuation setting.
Query Syntax
:CHANnel<n>:OFFSet?
The :CHANnel<n>:OFFSet? query returns the current offset value for the selected
channel.
Return Format
<offset><NL>
<offset> ::= Vertical offset value in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 203
":CHANnel<n>:RANGe" on page 218
":CHANnel<n>:SCALe" on page 219
":CHANnel<n>:PROBe" on page 212

Table of Contents

Related product manuals