:CHANnel<n> Commands 10
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 211
:CHANnel<n>:OFFSet
(see page 798)
Command Syntax
:CHANnel<n>:OFFSet <offset> [<suffix>]
<offset> ::= Vertical offset value in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:OFFSet command sets the value that is represented at center
screen for the selected channel. The range of legal values varies with the value set
by the :CHANnel<n>:RANGe and :CHANnel<n>:SCALe commands. If you set the
offset to a value outside of the legal range, the offset value is automatically set to
the nearest legal value. Legal values are affected by the probe attenuation setting.
Query Syntax
:CHANnel<n>:OFFSet?
The :CHANnel<n>:OFFSet? query returns the current offset value for the selected
channel.
Return Format
<offset><NL>
<offset> ::= Vertical offset value in NR3 format
See Also • "Introduction to :CHANnel<n> Commands" on page 203
• ":CHANnel<n>:RANGe" on page 218
• ":CHANnel<n>:SCALe" on page 219
• ":CHANnel<n>:PROBe" on page 212

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals