:MEASure:VTIMe
(see page 798)
Query Syntax
:MEASure:VTIMe? <vtime_argument>[,<source>]
<vtime_argument> ::= time from trigger in seconds
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VTIMe? query returns the value at a specified time on the source
specified with :MEASure:SOURce. The specified time must be on the screen and is
referenced to the trigger event. If the optional source parameter is specified, the
current source is modified.
Return Format
<value><NL>
<value> ::= value at the specified time in NR3 format
See Also • "Introduction to :MEASure Commands" on page 350
• ":MEASure:SOURce" on page 380
• ":MEASure:TEDGe" on page 382
• ":MEASure:TVALue" on page 384
This query is not available if the source is FFT (Fast Fourier Transform).