EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
724 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
31 Obsolete and Discontinued Commands
:FUNCtion:SOURce
(see page 798)
Command Syntax
:FUNCtion:SOURce <value>
<value> ::= {CHANnel<n> | ADD | SUBTract | MULTiply}
<n> ::= 1 to (# analog channels) in NR1 format
The :FUNCtion:SOURce command is only used when an FFT (Fast Fourier
Transform) operation is selected (see the:FUNCtion:OPERation command for more
information about selecting an operation). The :FUNCtion:SOURce command
selects the source for function operations. Choose CHANnel<n>, or ADD, SUBT, or
MULT to specify the desired source for FFT operations specified by the
:FUNCtion:OPERation command.
Query Syntax
:FUNCtion:SOURce?
The :FUNCtion:SOURce? query returns the current source for function operations.
Return Format
<value><NL>
<value> ::= {CHAN<n> | ADD | SUBT | MULT}
<n> ::= 1 to (# analog channels) in NR1 format
See Also "Introduction to :FUNCtion Commands" on page 291
":FUNCtion:OPERation" on page 303
NOTE
The :FUNCtion:SOURce command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :FUNCtion:SOURce1 command (see page 308) instead.

Table of Contents

Related product manuals